High Voltage Test Solutions
Persistently developing technology, improving quality, management, and service standards

Oil Testing Instrument

LQHV-680S Three-Cup Transformer Oil BDV Tester | Automatic Dielectric Strength Tester

The LQHV-680S three-cup oil BDV tester automates dielectric strength tests per GB507-86 & DL/T 596-1996 standards. Features microcomputer control, triple-sample testing, and overvoltage protection.


  • US$1.00; Negotiated Price
  • Lead time: 30days

>>>WhatsApp Communication<<<

Product Details

Product Tags

LQHV-680S Three-Cup Transformer Oil BDV Tester

Fully automated dielectric strength testing for insulating oils with triple-sample efficiency

Product Parameters

Technical parameters

Booster capacity1.5kVA
Booster speed2.0kV /s, 2.5kv /s, 3.0kV /s, 3.5kV /s
error: 0.2 kV/s
The output voltage0 ~ 80kV (optional)
Voltage accuracy± (2% reading +2 words)
Power distortion rate<1%
Electrode gapstandard 2.5 mm
Number of test6 times (1-6 times optional)
Let stand time5 min (1-9min optional)
Dimensions670mm×420mm×410mm
Weight38kg

Conditions of Use

Environment temperature0~40℃
Relative humidity≤85%
Working power supplyAC220V±20%
Power frequency50Hz(1±10%)
Power<200W

Technical index

Step-up transformer capacity1.5kVA
Rate of voltage rise2.0kV/s,2.5kV/s,3.0kV/s,3.5kV/s five gears
Output voltage0~80kV
Aberration rate of power supply<1%
Display modeBig LCD display Chinese characters
Electrode gapStandard 2.5mm
Boundary dimension730mm×410mm×390mm
Instrument weight38kg(ODST-1203)

Compliant with: GB507-86 | DL-474.4-92 | DL/T 596-1996

Why Choose the 680S Model?

  • Triple-Cup Efficiency – Test 3 samples sequentially without manual intervention

  • Military-Grade Protection – Overcurrent/voltage/short-circuit safeguards

  • Dual Standard Modes – Supports both GB1986 & GB2002 test protocols

  • Portable Rigidity – 10kg ruggedized design for lab/field use

Automated Workflow

  1. Sample Loading – Fill all 3 cups simultaneously

  2. One-Touch Start – Initiates:

    • Individual cup stirring (1500 RPM)

    • Voltage ramp-up (2kV/s)

    • Breakdown detection & recording

  3. Smart Analysis – Calculates:

    • Average breakdown voltage per cup

    • Inter-cup variance (

      <5% ideal="">

Fail-Safe Protections

  • 🛡️ Watchdog Circuit – Prevents system freezes

  • 🛡️ Auto-Zeroing – Resets between tests

  • 🛡️ Instant Discharge – Capacitor safety dump

Applications

Power Transformers

  • Factory acceptance testing (FAT)

  • Periodic maintenance checks

Oil Regeneration

  • Filtration efficiency validation

  • Reclaimed oil quality control

Advantages vs Single-Cup Testers

MetricLQHV-680S (3-Cup)Standard 1-Cup Testers
Testing Throughput3× fasterBase rate
Data ConsistencyCross-cup verificationSingle-point data

Maintenance Protocol

  • 🔧 Daily: Electrode cleaning with lint-free cloth

  • 🔧 Monthly: Calibration check with reference oil

  • 🔧 Yearly: Full factory recalibration

>>>WhatsApp Communication<<<

>>>Click To Get Quote<<<

Leave Your Message

Leave Your Message