Oil Testing Instrument
LQHV-660 Electrolytic Moisture Analyzer
ARM-powered Karl Fischer coulometric titrator for ppm-level water detection in insulating oils and industrial fluids
Compliant with: ASTM D6304 | IEC 60814 | GB/T 7600
Technical parameters
Titration method | Coulometric titration(coulometric analysis) |
Measurement range | 1ug~100mg(Typical value 10ug~100ug) |
Electrolytic current | 0-400mA |
Sensitivity threshold | 0.1ug |
Accuracy | 5ugH2O~1mgH2O ± 2ug, 0.3% above 1mgH2O (excluding injection error and environmental humidity error) |
Sample type | Solid, liquid, and gaseous states |
Display method | 64K color high-definition touch display |
Data Storage | 1000 test records |
Status indicator | Dynamic curves, text display |
Mixing speed regulation | Sliding touch panel speed control |
Printer | Micro thermal printer, paper width 56mm |
Power supply | AC220V±10V、50Hz±2.5Hz |
Power | 50VA |
Operating ambient temperature | 5~35℃ |
Operating environment humidity | ≤85% |
Overall dimensions | 330mm x 260mm x 220mm (length x width x height) |
Core Capabilities
💧 Ultra-Sensitive Detection - Measures 1ppm to 5% water content
📈 Real-Time Visualization - Dynamic potential curve monitoring
🧠 Smart Algorithms - Auto-calculation for weight/volume measurements
🔋 Self-Diagnostics - Electrode/reagent status monitoring
Testing Methodology
Karl Fischer Coulometry
Iodine Generation - Electrolytic cell produces I2 from reagent
Reaction:
H2O + I2 + SO2 + 3RN → 2RN·HI + RN·SO3
Endpoint Detection - Bivoltometric electrode senses excess I2
Calculation:
H2O (μg) = (Q × Mw) / (2 × F)
Q=Charge (Coulombs), Mw=18g/mol, F=Faraday constant
Smart Features
Dynamic Monitoring
Real-time titration curve display
Auto-zoom for endpoint identification
Adaptive Testing
Delay start for low-moisture samples
Adjustable stirring speed (50-1000 RPM)
Industry Applications
Industry | Typical Samples | Critical Moisture Levels |
---|---|---|
Power | Transformer oil | ≤35ppm (IEC 60422) |
Petrochemical | Lubricants | ≤50ppm (ASTM D4377) |
Pharma | APIs | ≤0.5% (USP<921>) |
Maintenance Protocol
🧪 Daily - Anode/cathode cleaning
🔄 Weekly - Reagent replacement (typically 50 tests/bottle)
📊 Monthly - Verify with sodium tartrate dihydrate