Oil Testing Instrument
LQHV-655 Open Cup Flash Point & Ignition Point Tester
ARM-controlled dual-test analyzer for petroleum product flammability assessment
Product Parameters
Technical Parameters
Measurement range | Room temperature~400 ℃ |
Detection type | Open flash point or ignition point |
Temperature detection | platinum |
Accuracy | ±1℃ |
Repeatability | conform to GB3536-2008(ASTM D92)、GB/T 267-88 |
Display | Color LCD display |
Information storage | Can store 1000 measurement results |
Ignition method | electric ignition |
Cooling method | Forced air cooling |
Printer | Dot matrix thermal printer |
Self inspection function | Lifting rod, sweeping rod, printing, etc |
Power | ≤600VA |
Using the power supply | AC 220V ± 11V, frequency 50Hz ± 2.5Hz |
Ambient temperature | 10~35℃ |
Ambient humidity | ≤90% |
Weight | About 17kg |
Complies with: ASTM D92 | GB3536-2008 | GB/T 267-88
Key Advantages
🔥 Dual-Function Testing - Measures both flash and ignition points in one cycle
🎯 Precision Heating - ±1.5°C temperature control (0.1°C resolution)
💾 Extended Memory - Stores 1,000+ test records with timestamps
⚡ Electric Ignition - No combustible gas required (enhanced lab safety)
Testing Process
Sample Preparation - 70±5ml oil poured into brass test cup
Test Initiation:
Electric igniter positioned 5mm above cup rim
PID-controlled heating ramp
Dual Detection:
Flash Point - First flame appearance (≥5mm height)
Ignition Point - Sustained burning (≥5 seconds)
Auto Shutdown - Cooling fan activates post-test
Safety Systems
🛡️ Over-Temp Protection - Cuts power at 410°C
🛡️ Spill Containment - Built-in fire-resistant tray
🛡️ Visual Alerts - Flashing LED during ignition cycles
Typical Test Results
Sample | Flash Point | Ignition Point |
---|---|---|
Transformer Oil (Mineral) | 145-160°C | 165-185°C |
Diesel Fuel | 52-96°C | 60-110°C |
Why This Tester?
Feature | LQHV-655 | Manual Methods |
---|---|---|
Test Time | 20-25 minutes | 60+ minutes |
Data Traceability | Digital records | Paper logs |
Maintenance Protocol
🧽 Post-Test - Clean cup with toluene/acetone mix
📅 Monthly - Verify with certified reference oils
🔍 Annually - Calibrate temperature sensors