Oil Testing Instrument
LQHV-665 Automatic Acid Value Tester
ARM-powered field analyzer for detecting oil oxidation in power equipment and industrial lubricants
Technical Parameters
Display | High definition color touch screen |
Accuracy | 0.01mgKOH/g between 0.001 and 0.1 |
0.03mgKOH/g between 0.1~0.5 | |
0.07mgKOH/g between 0.5 and 1.0 | |
0.10mgKOH/g between 1.0 and 2.0 | |
Resolution | 0.0001 mgKOH/g |
Measuring range | 0.0001~10.000 mgKOH/g |
Detection | Neutralization method |
Number of samples | three |
Supply voltage | AC220V ±10% 50HZ |
Operating ambient temperature | 5 ℃~50℃ |
Operating environment humidity | ≤75% |
External dimensions | L×B×H 440×280×200mm |
Printing | Thermal printing (40 line characters) |
Power | 60W |
Weight | 10kg |
Compliant with: ASTM D664 | IEC 62021-1 | GB/T 7304
Core Advantages
🧪 Contactless Operation - Sealed reagent system prevents operator exposure
📶 Wireless Control - Bluetooth/WiFi for remote monitoring
⚡ High Throughput - Processes 6 samples sequentially
🌍 Field-Ready - IP54 rated for harsh environments
Testing Principle
Potentiometric Titration
Sample Preparation - 5±0.1g oil + 50ml toluene/isopropanol
Auto-Titration - KOH solution (0.1mol/L) added incrementally
Endpoint Detection - pH electrode identifies neutralization point
Calculation:
Acid Value = (V × C × 56.1) / m
V=KOH volume(ml), C=concentration(mol/L), m=sample mass(g)
Smart Features
EMC Robustness
Meets IEC 801-4 (±4kV ESD protection)
Filters power grid fluctuations (±15%)
Safety Systems
CO2-blocking reagent bottles
Waste liquid sealed collection
Critical Oil Analysis
Oil Type | New Oil Limit | Condemnation Limit |
---|---|---|
Transformer Oil | ≤0.03 mg KOH/g | ≥0.1 mg KOH/g (IEC 60422) |
Turbine Oil | ≤0.2 mg KOH/g | ≥0.5 mg KOH/g (ASTM D4378) |
Operational Guide
🔄 Reagent Replacement - Every 100 tests (indicator light alert)
🧹 Electrode Care - Monthly polishing with alumina slurry
📱 Mobile App - View historical trends via Android/iOS