<100kV Range
LQHVF 50kV/1000pF Capacitive Voltage Divider & Coupling Capacitor
Compact Powerhouse: Dual-function HV instrument in space-efficient 300×300×600mm footprint with full 50kV capability
Precision HV Instrument for Demanding Applications
Engineered for 50kV systems, the LQHVF combines laboratory-grade accuracy with field-deployable convenience. The 1000pF capacitance and 1000:1 ratio enable:
Accurate voltage measurement with ≤±1% accuracy
Sensitive partial discharge detection (<5pC)
Harmonic analysis up to 20kHz bandwidth
Dielectric loss monitoring with <0.5% dissipation
Technical Specifications
Parameter | Specification | Performance Advantage |
---|---|---|
Rated Voltage | 50kV | Ultra-high voltage capability |
Capacitance | 1000pF | Optimal PD signal coupling |
Voltage Ratio | 1000:1 | 50V output at 50kV input |
Partial Discharge | <5pC @ 50kV | IEC 60270 Class I compliance |
Dielectric Loss | <0.5% | Superior insulation efficiency |
Accuracy | ≤±1% | Meets IEEE 4-2013 standards |
Dimensions (L×W×H) | 300×300×600mm | Compact substation footprint |
Weight | 15kg | Portable with carrying handles |
Operating Time | 60min @ 50kV | Full-power diagnostic capability |
L300mm
W300mm
H600mm
Innovative Modular Design
Corona Ring
Aluminum alloy
Removable design
Ly Arm
Measurement interface
BNC output
HV Arm
50kV rated
Quick-disconnect
Removable Base
Stabilized platform
Leveling feet
Oil-Immersed Construction
High-grade mineral oil for thermal management
Benefit: Enhanced arc suppression
Quick-Disconnect System
Tool-free assembly in 8 minutes
Benefit: Rapid field deployment
Dual-Function Capabilities
Voltage Division
Range: 0-50kV
Accuracy: ≤±1% at 50Hz
Output: 50V at 50kV input
PD Coupling
Sensitivity: <5pC at 50kV
Bandwidth: 20Hz-20kHz
Impedance: 80Ω at 2kHz
Measurement Advantage: Simultaneous voltage and PD measurement enables phase-resolved partial discharge (PRPD) analysis
Thermal Management System
ONAN Cooling: Oil Natural Air Natural convection
Thermal Performance: <45°C rise during 60-min operation
Operating Protocol:
60 minutes continuous at 50kV
Recommended 60-minute cooling period
Clearance Requirements: Maintain 25cm air space around unit
Compliance & Applications
Standards Compliance: IEC 60270:2000, IEEE 4-2013, EN 60060-1:2010
Critical Applications:
Testing of 50kV switchgear and GIS
On-line PD monitoring in substations
Power quality analysis in industrial facilities
Dielectric testing of transformers
High-voltage research laboratories
Operational Protocol
Assembly
Secure base to stable surface
Connect Ly-arm to base
Install HV-arm assembly
Attach corona ring
Safety Measures
Ground base with ≥12mm² cable
Maintain 1.5m personnel clearance
Verify oil level before energization
Operation
Ramp voltage gradually (≤5kV/s)
Max 60 minutes at 50kV
Monitor temperature
Measurement
Connect Ly-arm to voltmeter
For PD: Connect HV terminal to analyzer
Record data continuously
Pro Tip: For optimal PD sensitivity, use double-shielded coaxial cables ≤1.5m in length with ferrite cores at both ends
Why This Compact Design?
Space Efficiency
300×300×600mm fits confined spaces
Transport Optimized
15kg with removable components
Full Capability
50kV performance in compact form
Thermal Resilience
ONAN cooling maintains stability