<100kV Range
LQHVF 20kV/500pF Capacitive Voltage Divider & Coupling Capacitor
Dual-Function Design: Precision Voltage Measurement + PD Detection
Industrial-Grade HV Instrument for Critical Systems
Engineered for 20kV applications, this dual-function device combines a 1000:1 voltage divider with PD coupling capabilities. The 500pF capacitance delivers:
Accurate RMS voltage measurement with ≤±1% accuracy
High-sensitivity partial discharge detection
Harmonic analysis up to 20kHz bandwidth
Dielectric loss monitoring in transformers and cables
Technical Specifications
Parameter | Specification | Performance Advantage |
---|---|---|
Rated Voltage | 20kV | Industrial power system compatibility |
Capacitance | 500pF | Optimal balance for measurement & coupling |
Voltage Ratio | 1000:1 | Direct interface with standard instruments |
Dielectric Loss | <0.5% | Superior insulation efficiency |
Accuracy | ≤±1% | Meets IEEE 4-2013 Class 1 standards |
Dimensions | 300×400×400mm | Compact substation footprint |
Weight | <50kg | Portable with lifting handles |
Thermal-Managed Operation Cycle
Operation Phase
Duration: 60 minutes
Conditions: 100% rated voltage
Max ΔT: 65K (winding)
Cooling Phase
Duration: 60 minutes
Cooling: Natural convection
Max ΔT: 55K (winding-oil)
Daily Capacity
Cycles: 4 full operations
Total Runtime: 4 hours
Designed for 8-hour shifts
Thermal Advantage: Oil-immersed design maintains safe operating temperatures during intensive testing schedules.
Modular Structural Design
Top Corona Cover
Aluminum alloy
Prevents surface discharge
HV Arm
Removable connection
20kV rated
L Arm
Measurement interface
BNC output
Stabilized Base
Non-slip design
Leveling feet
Oil-Immersed Construction: Mineral oil enhances heat dissipation and arc suppression
Quick-Disconnect: Field assembly in under 10 minutes
Corona-Free Design: Optimized geometry prevents RF interference
PD Measurement Capability
Compliance: Meets IEC 60270 Partial Discharge Standards
Coupling Configuration:
Connect HV terminal to test object
Route L-arm output to PD analyzer
Ground base to earth reference
Use 50Ω coaxial cable ≤2m length
Detection Advantage: 500pF capacitance provides optimal impedance for PD signal coupling in 20kV systems.
Compliance & Applications
Standards Compliance:
IEC 60270:2000 (Partial Discharge Measurements)
IEEE 4-2013 (High-Voltage Testing)
EN 60076-11 (Dry-Type Transformers)
Critical Applications:
Factory acceptance testing of 20kV switchgear
On-line PD monitoring in industrial facilities
Dielectric loss measurement in power transformers
High-voltage research laboratories
Operational Protocol
Assembly: Connect modules in sequence (base → L-arm → HV arm → corona cover)
Grounding: Establish earth connection with ≥10mm² cable
Cooling Check: Verify oil level before energization
Operation Cycle:
Max 60 minutes continuous at 20kV
Follow with 60-minute cooling period
Max 4 cycles per 8-hour shift
PD Measurement: Use shielded coaxial cable directly to analyzer
Safety Feature: Automatic discharge circuit activates when voltage drops below 50V