<150kV Range
LQHDR Series 12kV/1kΩ Compact Damping Resistor
Space-Efficient HV Damping Solution
Designed for compact test systems, the LQHDR damping resistor delivers <5pC partial discharge performance at 12kV in an ultra-compact Ø70×115mm package. With 1kΩ resistance and flexible operating modes, it provides precision damping for:
Portable PD test sets
Transformer diagnostic equipment
Field testing instruments
Lab-grade HV measurement
⌀70mm
↓
115mm
Technical Specifications
Parameter | Specification | Standard |
---|---|---|
Model | LQHDR | - |
Rated Voltage | 12 kV AC/DC | IEC 60137 |
Nominal Resistance | 1 kΩ ±1% | IEC 60255 |
Rated Current | 0.1 A | IEC 60076 |
Power Rating | 10W @ full load | - |
Partial Discharge (PD) | <5 pC at 12kV | IEC 60270 |
Dimensions | Ø70 × 115mm | - |
Weight | <5 kg | - |
Terminals | BNC connectors | IEC 61169-8 |
Flexible Operating Modes
Standard Operation
60 minutes continuous at 100% rated voltage (12kV) and current (0.1A)
Extended Operation
Unlimited runtime at 80% rated voltage (9.6kV) and current (0.08A)
Thermal Management
Maintains surface temperature rise ≤80K during continuous operation at 100% load through:
Aluminum oxide ceramic core with high thermal conductivity
Optimized surface-area-to-volume ratio
Passive cooling fins integrated into housing
Key Features
Ultra-Compact Design
At just Ø70×115mm and <5kg, this is one of the smallest 12kV damping resistors available - ideal for portable test equipment and confined spaces.
Laboratory-Grade Precision
1kΩ ±1% tolerance ensures accurate waveform control in:
HV probe calibration
Divider ratio verification
Oscillation damping circuits
Low-PD Construction
Triple-shielded design maintains <5pC background noise for sensitive partial discharge measurements per IEC 60270.
Field Applications
Portable PD Testing
Critical component in handheld partial discharge detectors for:
Substation equipment checks
Cable termination assessments
Rotating machine diagnostics
Transformer Testing
Provides stable damping for:
Turns ratio test sets
Winding resistance measurements
Impulse testing circuits
Installation & Safety
Mounting: Vertical or horizontal orientation
Clearance: Minimum 30mm to grounded surfaces
Connections: BNC connectors (50Ω coaxial)
Safety: Automatic discharge circuit (RC time constant <5s)
Environment: -25°C to +55°C operating range