LQHDR 1000kV/5kΩ/0.67A Damping Resistor: ONAN-Cooled Cyclic Operation

Technical specifications for the LQHDR 1000kV/5kΩ/0.67A damping resistor featuring ONAN cooling, cyclic operation (15min on/off), and <5pC partial discharge performance.


  • Negotiated Price
  • Lead time: 60days

>>>WhatsApp Communication<<<

Product Details

Product Tags

LQHDR 1000kV/5kΩ/0.67A ONAN-Cooled Damping Resistor

Electrical Specifications

ParameterSpecificationStandard
Rated Voltage1000kV ACIEC 60273
Nominal Resistance5kΩ ±0.3%IEEE 4-2013
Rated Current0.67AIEC 60076
Power Rating2.2445kWThermal Class A
Frequency50HzIEC 60060-1

Partial Discharge Performance

  • <5pC at 100% rated voltage (1000kV)

  • <3pC at 80% rated voltage (800kV)

  • Measured per IEC 60270 Class I precision

Ultra-low discharge characteristics ensure measurement integrity in sensitive high-voltage circuits.

Cyclic Operation Capability

Duty Cycle: 15 minutes ON / 15 minutes OFF

Maximum Cycles: 8 per 24-hour period

Optimized for intermittent high-stress test sequences in laboratory environments.

ONAN Cooling System

  • Cooling Method: Oil Natural Air Natural (ONAN)

  • Oil Capacity: 220L mineral oil (IEC 60296)

  • Thermal Capacity: 2.5kW heat dissipation

  • Cooling Surface: 8m² corrugated radiator

The passive cooling system ensures zero-maintenance operation with natural convection.

Thermal Performance

  • Stabilizes within 15 minutes of operation

  • 15-minute cooldown period between cycles

  • Oil temperature rise maintained below 55K

Cyclic Duty Applications

  • Intermittent transformer impulse testing

  • Batch testing in high-voltage laboratories

  • Research facility validation cycles

  • Quality assurance sampling systems

  • Prototype evaluation test sequences

Operational Protocol

  1. Pre-cycle insulation test (min 50GΩ at 5kV DC)

  2. Maximum 8 cycles per 24-hour period

  3. 15-minute minimum cooldown between operations

  4. Annual oil dielectric strength verification

Safety & Reliability Features

  • Pressure-relief membrane for oil expansion

  • Corona-free design up to 1100kV

  • Dual moisture sensors in oil compartment

  • Grounding terminals with 10mm² minimum requirement

Compliance Standards

  • IEC 60273: High-voltage resistors

  • IEEE 4-2013: HV testing standards

  • EN 60076-14: Liquid-immersed equipment

  • IEC 61010-1: Safety requirements