≥1000KV Range
LQHCG 3600KV-400pF Ultra-High Voltage Chopping Gap System
Record-Breaking High Voltage Testing
Pushing the boundaries of high-voltage testing, the LQHCG 3600KV-400pF delivers unprecedented 3600kV capacity with laboratory-grade ≤0.1μs chopping precision. This system sets new global benchmarks for UHV equipment testing and high-energy physics research.
Technical Specifications
Core Parameters
Rated Voltage: 3600kV (peak)
Rated Capacitance: 400pF ±0.8%
Energy Storage: 2.59MJ at full voltage
Chopping Range: 2-6μS adjustable
Precision Control
Trigger System: 3-electrode discharge
Trigger Pulse: 10kV, 100ns rise time
Time-Delay Pulse: 250kV adjustable
Chopping Precision: σ<0.1μs
Mechanical Design
Sphere Gaps: 8 pairs (4 per stage)
Gap Diameter: 150mm semi-sphere
Clearance: Minimum 8m at 3600kV
Safety Systems: Automatic discharge
Breakthrough Technology
World-Record Voltage
3600kV capacity enables testing of next-generation 1200kV+ UHV transmission equipment
Nanosecond Triggering
10kV/100ns pulse amplifier ensures reliable initiation at extreme voltages
Unmatched Precision
Sub-0.1μs chopping time deviation meets metrology lab standards
Performance Comparison
| Parameter | LQHCG 3600KV | Industry Standard |
|---|---|---|
| Maximum Voltage | 3600kV | 2400-3000kV |
| Trigger Speed | 100ns | 200-500ns |
| Chopping Precision | σ<0.1μs | σ<0.3-0.5μs |
| Energy Capacity | 2.59MJ | 1.0-1.8MJ |
Standards Compliance
Certified to all major international standards:
IEC 60060-1:2023 High-voltage test techniques
IEEE Std 4-2023 Standard for HV testing
GB/T 16927.1-2023 Lightning impulse tests
EN 60060-2:2023 Measurement requirements
Typical Applications
1200kV+ UHV transformer testing
Next-generation HVDC equipment validation
National high-voltage reference standards
High-energy physics research
Nuclear fusion facility diagnostics
Safety Protocol
Operational Requirements
8m minimum clearance during 3600kV operation
Daily inspection of all sphere gaps
Remote operation mandatory above 3000kV
Maintenance Schedule
| Interval | Action | Tolerance |
|---|---|---|
| Every 25 operations | Sphere gap alignment | ±0.05mm |
| Monthly | Trigger circuit verification | 10kV±2% |
| Annual | Full system recalibration | IEC 60060-3 Class 0.5 |
Why This System?
The LQHCG 3600KV-400pF represents the ultimate achievement in high-voltage testing technology, delivering world-record 3600kV capacity with metrology-grade ≤0.1μs precision. This system is engineered for national research laboratories and power equipment manufacturers pushing the boundaries of ultra-high voltage technology.