150kV ~ 400kV
LQHDR 200kV/5kΩ Damping Resistor: Ultra-Low Partial Discharge Performance
Precision Component for Critical HV Measurements
The LQHDR 200kV/5kΩ damping resistor sets new standards in high-voltage testing with its exceptionally low partial discharge (PD) characteristics. Engineered for 50Hz systems at 200kV rated voltage, this specialized component minimizes electrical noise during sensitive dielectric tests while maintaining thermal stability under continuous 1-hour operation. Essential for utilities, certification labs, and transformer manufacturers requiring uncompromised measurement accuracy.
Technical Specifications
Parameter | Specification |
---|---|
Model | LQHDR 200kV/5kΩ/0.1A |
Rated Frequency | 50Hz |
Rated Voltage | 200kV |
Rated Current | 0.1A |
Nominal Resistance | 5kΩ |
Rated Power | 50W* |
Partial Discharge Level | ≤2pC at rated voltage |
Operating Time | 1 hour continuous |
Temperature Rise | ≤80K at 100% load |
*Calculated: P = I²R = (0.1A)² × 5kΩ = 50W
Performance Advantages
Ultra-Low Noise Measurement
With PD levels ≤2pC at 200kV - 5x lower than industry standards - this resistor eliminates measurement interference during precision dielectric testing. Exceeds IEC 60270 requirements for partial discharge detection in transformers and cables.
Thermal Reliability
Maintains ≤80K surface temperature rise during full-load operation, preventing resistance drift. Proprietary cooling design ensures stable performance throughout 1-hour test cycles.
High-Voltage Integrity
Optimized for 50Hz systems with corona-free construction. Voltage grading technology prevents surface tracking even in high-humidity environments.
Key Applications
Transformer PD Testing – Critical for IEC 60076-27 compliance measurements
HV Cable Diagnostics – Dampens reflections in TDR/VLF test systems
Circuit Breaker Testing – Provides stable load for timing and contact resistance tests
Research & Certification Labs – Reference component for calibration of PD measurement systems
Installation & Safety Guidelines
Mounting Requirements
Minimum clearance: 1.5m to grounded surfaces
Use silicone-insulated HV leads (≥200kV rating)
Grounding terminal must be bonded to test bench earth
Operational Precautions
Pre-heat for 15 minutes at 10% rated voltage before full-load tests
Monitor temperature with IR thermometer during extended operation
Allow 2-hour cooldown between test cycles
Complies with IEC 61010-1 safety standards for HV test equipment.
Why Choose LQHDR for Precision HV Testing?
Compared to Standard Resistors
▶ 80% lower PD noise than conventional wirewound resistors
▶ 40% faster thermal recovery between tests
▶ Validated for calibration labs (ISO 17025 traceable)