≤100kV Range
Technical Specifications
Electrical Performance
Parameter | Specification | Standard |
---|---|---|
Rated Capacity | 15 kVA | IEC 60076-11 |
Output Voltage | 100 kV AC | |
Output Current | 150 mA | |
Input Voltage | 0-400 V (Customizable) | |
Frequency | 50 Hz | IEC 60076-1 |
No-Load Current | ≤10% | IEEE C57.12.00 |
THD | ≤3% | IEC 61000-3-6 |
Dissipation Factor | <0.5% | IEC 60247 |
Measuring Voltage | 100 V |
Critical Performance Metrics
Partial Discharge: ≤2 pC at 100% rated voltage
Dielectric Strength: 5 kV/min (LV input to ground) - 40% higher than standard
Noise Level: <60 dB(A) at 1m distance
Thermal & Mechanical
Cooling System: ONAN (Oil-Natural Air-Natural)
Dimensions: 700 × 700 × 1250 mm (L×W×H)
Weight: 210 kg
Protection Class: IP54 (IEC 60529)
Key Advantages
✅ Lab-Grade Precision: ≤2pC PD for critical insulation diagnostics
✅ Enhanced Safety: Industry-leading 5kV/min dielectric strength
✅ Silent Operation: <60dB noise ideal for lab environments
✅ High-Efficiency Design: ≤3% THD ensures pure output waveform
✅ Universal Input: 0-400V customizable primary voltage