Load Capacitor
HRC 100kV/30,000pF Reference Capacitor
Primary Application: Calibration and precision measurements in HV test systems
Electrical Specifications
Parameter | Specification | Standard |
---|---|---|
Rated Voltage | 100 kVrms @60Hz | IEC 60358-1 |
Withstand Voltage | 110 kV/1min | IEC 60060-1 |
Capacitance | 30,000 pF ±1% | IEC 60384-1 |
Partial Discharge | <2 pC @100kV | IEC 60270 |
Precision Characteristics
Measurement Accuracy
Dielectric Loss: ≤0.2% @20°C
Voltage Ratio Error: ≤±1%
Phase Angle Error: ≤0.02°
Thermal Performance
Cooling: ONAN (Oil Natural Air Natural)
Temp Coefficient: <30 ppm/°C
Operating Range: -10°C to +50°C
Mechanical Design
Insulation: Epoxy-oilpaper composite tube
Wall thickness: 25mm minimum
3-layer graded shielding
Base:
4× locking casters with ±3° leveling
Quick-release mounting system
Weight: ≈320 kg (operational)
HV Terminal: Corona-free design (Ø500mm ring)
Installation Requirements
Clearance:
3m minimum to grounded surfaces
2m vertical clearance for lead connection
Measurement Setup:
Use supplied 25kV-rated triaxial cable
Shield termination within 0.5m of source
Pre-Test: Energize at 30kV for 15min before full voltage
Calibration & Maintenance
Interval | Procedure |
---|---|
Every 500h | Dielectric loss measurement @50kV |
Annual | Oil sample analysis (IEC 60422) |
After transport | Verify base level and grounding |