LQ-905 Intelligent CT/PT Calibrator - Precision Transformer Testing Instrument
Product Overview
The LQ-905 Intelligent CT/PT Calibrator is an advanced testing instrument designed for precision measurement of transformer ratio errors, polarity verification, and impedance admittance in current and potential transformers. This professional-grade equipment combines DSP technology with intuitive operation for reliable field and lab testing.
Technical parameters
Equipment operating environment | temperature:5℃~40℃ | relative humidity:<80%(25℃时) |
altitude:<2500m | ||
mains frequency:50Hz±0.5Hz | mains input:220V±5V | |
range of measurement | cophase component(%):0.0001~200.0 | distinguishability:0.0001 |
orthogonal component(part):0.001~700.0 | distinguishability:0.001 | |
impedance(Ω):0.0001~20.0 | distinguishability:0.0001 | |
admittance(ms):0.0001~20.0 | distinguishability:0.0001 | |
intrinsic error | cophase component | ΔX=±(X×2%+Y×2%)±Dx(Level 1 is optional) |
orthogonal component | ΔY=±(X×2%+Y×2%)±Dy(Level 1 is optional) | |
"X"、"Y"——Display value of instrument | ||
"Dx,Dy"——Display value of instrument | ||
Dx=2,Dy=5 | ||
dial indicator | PETS2(Level 1 is optional) | |
working range | electricity | (1%~149%)In (In =5A) |
(5%~149%)In (In =1A) | ||
voltage | (5%~149%)Un (Un =100V,150V,100V/)![]() | |
(5%~149%)Un (Un =100V/3) | ||
workload | electricity | TO to TX<0.12Ω cosΦ=1 |
voltage | a to x<0.25VA (100V) | |
Polarity error indication | When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 180%, there should be polarity indication. | |
Note: if it is more than 10% of the rated operating current (voltage), there is still no proper polarity indication, indicating a fault. Please do not increase the current (voltage) to avoid burning out the instrument. | ||
Variable ratio error indication | When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 30% but is less than 180%, the error indication of variable ratio should be given. | |
Insulation and voltage test and instructions | terminalTX(![]() | |
The power socket can withstand the housing1.5kV,1minwithstand voltage |
Key Features
High-Resolution Display
320×240 lattice LCD with long-life backlighting for clear visibility in all lighting conditions
Automatic Diagnostics
Automatically detects polarity errors and calculates ratio errors with precision
Measurement Excellence
Solves S-class measurement challenges with automatic range switching capability
Regulatory Compliance
Fully meets JJG314-1993 and JJG314-1994 standards requirements
Advanced Technology
DSP-Enhanced Circuitry: Combines advanced circuit design with digital signal processing for stable measurements
Virtual Instrument Function: Upper computer software expands testing capabilities
Automatic Range Selection: Eliminates manual switching errors
S-Class Resolution: Handles precise measurement requirements for high-accuracy transformers
Typical Applications
Power Substations
Field verification of CT/PT transformer accuracy
Manufacturing QC
Production line testing of transformer components
Calibration Labs
Reference standard for instrument calibration
Maintenance Teams
Periodic verification of installed transformers
Operation Highlights
User-friendly Chinese interface with intuitive navigation
Large-font display of percentage difference and angle difference
Automatic sampling function reduces operator intervention
Micro printer compatibility for report generation