HET 25kVA High-Voltage Exciter Transformer (12.5kV/6.25kV) – Technical Datasheet

Precision 25kVA exciter transformer with 12.5kV/6.25kV dual outputs, ultra-low partial discharge (<2pC), and electrostatic shielding. Ideal for high-voltage testing labs and industrial applications.


  • US$ 4,672.00
  • Lead time: 30 day

>>>WhatsApp Communication<<<

Product Details

Product Tags

HET 25kVA High-Voltage Exciter Transformer (12.5kV/6.25kV) – Technical Datasheet

Product Overview

The HET 25kVA exciter transformer is engineered for precision high-voltage testing with dual-range outputs (12.5kV/6.25kV). Featuring ultra-low partial discharge (<2pC) and advanced electrostatic shielding, it complies with IEC 60076-11 for dielectric testing in high-voltage laboratories and industrial facilities.

Electrical Specifications

ParameterSpecification
Rated Capacity25kVA
Input Voltage0.4kV
Input Current62.5A
Output Configurations:
• High Range12.5kV @ 2A
• Standard Range6.25kV @ 4A
Frequency50Hz ±0.5%
Impedance Voltage<4%
Partial Discharge<2pC at rated voltage (IEC 60270)

Advanced Protection Systems

1. Overvoltage Protection

  • Piezoresistors on all HV outputs

  • Zinc oxide arresters (IEC 60099-4 compliant)

  • Ground-connected via 15kA surge arresters

2. Insulation & Discharge Control

  • 12.5kV corona-free HV bushings

  • Electrostatic shield layer between HV windings

  • Isovolumetric capacitance cores for stable voltage gradient

Mechanical Construction

Enclosure Features

  • Steel housing with epoxy powder coating

  • Wheel-mounted (4× heavy-duty casters)

  • Shell-type oil tank with 40L conservator

Terminal Configuration

  • LV terminals: Porcelain bushings

  • HV terminals: Fiberglass-reinforced bushings

  • Dual corona rings per HV bushing

Winding Technology

Advanced Insulation

  • Cylinder-type winding with balanced top/bottom capacitance

  • Inter-turn insulation (1.2mm Nomex® layers)

  • Electrostatic compensation shields for uniform voltage distribution

Dielectric Performance

  • Impulse withstand: 150kV (1.2/50μs wave)

  • Power frequency withstand: 30kV/min

  • Partial discharge <2pC at 1.1Ur

Installation & Maintenance

Commissioning Requirements

  • Minimum 1.5m clearance from grounded objects

  • Oil dielectric strength ≥55kV before energization

  • Initial insulation resistance ≥12,000MΩ

Service Schedule

  • Semi-annual oil testing (BDV, moisture content)

  • Bushing inspection every 3 months

  • Winding resistance test every 5,000 operating hours

Typical Applications

  • Power transformer factory acceptance tests

  • GIS (Gas Insulated Switchgear) commissioning

  • High-voltage cable partial discharge measurement

  • Rotating machine insulation diagnostics